Connect Vista with EZMQC

Follow

EZMQC Version 4.88 and above have ability to connect with current Vista Sensors. For Vista with serial number less than VTS00135 probably need to have some hardware added and the software updated (Vista Essentials 1.00.14 and above).

 

Connect Vista to Network through an Ethernet port

 1. Both of Vista and PC with EZMQC must be connected to Ethernet ports with same Ethernet network. 

 

 

 

 2. Open EZMQC in the PC. In Sensor, click "Add Sensor" and select "Vista". Select “Ethernet” and check the box "Discover and Select a Sensor in the Network" and then click "Search" to do an automatically searching. There will be a drop-down list of all available Vista sensors. If the sensor in the list including IP address as well as sensor name, then it is connectable. If the sensor in the list with “??????” instead of sensor name, it means that EZMQC can find the VISTA, while VISTA is not free to connect to EZMQC. If you meet this problem, you can restart VISTA and click search again. 

3. Click Next to do standardization.

 

Connect Vista to PC through an Ethernet cable directly 

1. Remove exiting Ethernet cable from PC. Connect Vista and PC with an Ethernet cable. Then restart Vista and PC. 

2. Open Command Prompt in the PC. Type in "ipconfig" and check the "autoconfiguration IPv4 Address" as well as the "Subnet Mask". 

 

3. Open Vista Essentials, go to Preferences/Configure Network Settings. Uncheck "Use DHCP for Ethernet Config". Type in IP address and Subnet Mask manually, then apply. The IP address here should be exact same as the "autoconfiguration IPv4 Address" in the PC, except changing the last two digits. Restart Vista to get network setting applied.

 

4. Open EZMQC in the PC, In Sensor, click "Add Sensor" and select "Vista". Select “Ethernet” and uncheck the box "Discover and Select a Sensor in the Network" and then type in the IP address which have been set up in Vista Essentials. 

5. Click Next to do standardization.

 

Connect Vista to PC through Wireless Access Point

1. About the connection and setup, please find the detail in this link, https://support.hunterlab.com/hc/en-us/articles/217231926-Connect-VISTA-through-Wireless-Access-Point-to-a-Computer-for-EZMQC-and-CalVer-


2. In Sensor, click "Add Sensor" and select "Vista". Select “Ethernet” and check the box "Discover and Select a Sensor in the Network" and then click "Search" to do an automatically searching. There will be a drop-down list of all available Vista sensors. If the sensor in the list including IP address as well as sensor name, then it is connectable. If the sensor in the list with “??????” instead of sensor name, it means that EZMQC can find the VISTA, while VISTA is not free to connect to EZMQC. If you meet this problem, you can restart VISTA and click search again. Please check the attached pictures. 


3. Click Next to do standardization.

 

 

 

 

Was this article helpful?
0 out of 0 found this helpful

Comments

Powered by Zendesk