What tolerance should be used for a PQ Performance Qualification on Transmission Color Measurements? Follow
FAQ: “What is the limit on the standardization in TTRAN. Using the small sample flask, filled with Ultra pure water. If they follow the protocol, they expect to find values (CIE Lab: 100.0; 0.0; 0.0)?”
When the instrument is standardized in TTRAN Total Transmittance mode, two levels are set:
- The Light Blocker serves as a physical reference for bottom-of-scale that sets 0% transmittance across the visible spectrum.
- The top-of-scale is a physical reference sets 100% transmittance across the visible spectrum. We also ask for the white standard tile to be placed at the reflectance port to fill in the sphere.
- If standardizing in TTRAN transmittance for transparent solids like plastic plaques, the top-of-scale standard is typically AIR. This is what is at the TTRAN port against the sphere.
- If standardizing in TTRAN transmittance for transparent liquids, the top-of-scale is typically a transmittance cell of some fixed path length filled with DI water.
As a PQ (Performance Qualification) check, independent of the instrument, we advise clients read back the top-of-scale standard as verification that the instrument is ready for measurement.
In transmittance standardization, the instrument is forced to the top-of-scale (air for solids; cell + DI water for liquids). If the operator reads the top-of-scale back, the measured values for 100% transmittance should match CIE L* = 100, a* = 0, b* = 0 (Yellowness Index = 0, APHA/Pt-Co = 0 if using) closely.
You raise the question of what is meant by “closely”. The metrics in our software are typically reported to 2 decimal places reflecting the degree of precision possible with the instrument. Ideally, on a PQ read back of top-of-scale, the measured values would be CIE L* = 100.00, a* = 0.00, b* = 0.00 but this requires some qualification.
- If you are standardizing on AIR as top-of-scale and using the largest area of sample view (LAV), this is what you would typically get on a PQ read back using a HunterLab sphere instrument, although a round off of +/- 0.01 is possible.
- If you are using SAV (Small Area of View) which has less optical area averaging of the sample view, you could be up to +/- 0.02 or 0.03.
- If you are standardizing on a cell filled with DI water, then the measurement is made through two glass or plastic windows of the cell, plus transmitted through DI water. This refraction can lead to some additional light bending on the PQ read back of cell + DI water versus AIR. Although typically it should be less, it is possible that an operator could see a PQ variation of as much as a 0.04.
- In addition, if the customer is using a micro-cell holder that has front and back optics in the holder that shrink the bean diameter to 2 mm, along with the effects of cell and DI water, this can lead to more light bending in the path independent of the sensor performance. For this situation, I typically suggest a PQ read back tolerance of +/- 0.3 in CIE L* = 100, a* = 0, b* = 0 target with L* typically showing the highest variation given the magnitude of the L* value.
Remember that this PQ verification step is not a performance test of the instrument – there are separate diagnostic tests of the sensor built into the software that can be run to verify instrument performance.
This PQ Performance read back of the top-of-scale standard is a test to verify that the operator has set up the instrument and standardized correctly in preparation for transmittance measurements. That is, the operator and instrument are “performance qualified” to continue to make sample measurements if they read back the top-of-scale transmittance standard “closely”.
As this test is outside HunterLab sensor performance, the customer can establish their own PQ tolerance based on the way they are making their transmittance measurement, or if they would like HunterLab to make a recommendation, +/- 0.3 in CIE L* = 100, a* = 0, b* = 0 should take care of most cases.
If the PQ read back is beyond that level, the operator should re-standardize and run the PQ test again.
Do you still need more information? Submit a ticket and our support team member will reach out to you soon!