How do you validate a APHA/Pt-Co/Hazen 500 liquid color standard? Follow
As described in Section 6 of ASTM D1209, there are 2 criteria to validate a APHA/Pt-Co 500 standard:
- The APHA/Pt-Co 500 Color Standard must have a defined chemical composition and be prepared in accordance with section 6 of ASTM D1209.
- After preparation, the APHA/Pt-Co 500 standard must meet a absorbance criteria at 4 points in the blue region of the visible spectrum per Table 1 of ASTM D1209.
- Platinum-Cobalt Reference Standards
6.1 Platinum-Cobalt Stock Solution— Dissolve 1.245 g of potassium chloroplatinate (K2PtCl6) and 1.00 g of cobalt chloride (CoCl2·6H2O) in water. Carefully add 100 mL of hydrochloric acid (HCl, sp gr 1.19) and dilute to 1 L with water. The absorbance of the 500 platinum-cobalt stock solution in a cell having a 10 mm light path, with reagent water in a matched cell as the reference solution, must fall within the limits given in Table 1.
ASTM D1209 TABLE 1 Absorbance Tolerance Limits For No. 500 Platinum-Cobalt Stock Solution
Wavelength, nm | Absorbance, A |
430 | 0.110 to 0.120 |
455 | 0.130 to 0.145 |
480 | 0.105 to 0.120 |
510 | 0.055 to 0.065 |
How to Validate a APHA/Pt-Co/Hazen 500 Color Standard
- In the Spectral Data View of EasyMatch QC, place your mouse over the view and Right Click > CONFIGURE to report Absorbance spectral data instead of Transmittance. If you do not have a Spectral Data View displayed, put your mouse over the COLOR DATA TABLE > Right Click > SPLIT Horizontally and Add SPECTRAL DATA VIEW.
- Fill a 10 mm path length cell with distilled water and place on transmittance bench on a HunterLab diffuse d/8 sphere instrument.
- Standardize in Total Transmittance (TTRAN), LAV (Large Area of View), UV Filter in the Nominal position using the Light Blocker and 10 mm cell + DI water to set bottom and top-of-scale.
- As a PQ Performance Qualification step, read back the cell + DI water as an OQ (Operation Qualification) step. The transmittance at 430, 455, 480 and 510-nm should be 100% with the corresponding absorbance being 0%.
- Read the APHA/Pt-Co 500 Liquid Color Standard to see if it reads within the absorbance tolerances at 430, 455, 480 and 510 nm in Table 1 of ASTM D1209. If your instrument only reports at 10 nm intervals. Average the absorbance at 450 and 460 nm.
If the APHA/Pt-Co/Hazen 500 Color Standard meets both the chemical and absorbance criteria of ASTM D1209, it is a 500 Standard.
Do you still need more information? Submit a ticket and our support team member will reach out to you soon!