How do I setup up the Color Data View in EasyMatch QC for APHA Color? Follow
FAQ: "One extra question - how do you set up EasyMatch QC to measure APHA Color values? It is measuring the Lab values right now."
In EasyMatch QC+ER, put your mouse over the COLOR DATA VIEW > Right Click > CONFIGURE, selecting the APHA metric from Indices on the right that matches your cell path length - APHA-10mm, APHA-20 mm, or APHA-50 mm.
Select SENSOR > SET MODE for Total Transmittance (TTRAN), Large Area of View (LAV), UltraViolet (UV) Filter Nominal and standardize using cell filled with distilled water as your top-of-scale set to 100% transmittance.
As a PQ (Performance Qualification) step, read back the cell + distilled water as a standard. It should read L* = 100, a* = 0, b* = 0, APHA-XXmm = 0 closely if everything is set up correctly.
If your measurement is performance qualified, proceed to measure the APHA Color of your transparent liquid samples.