D25LT calibrating touchscreen Follow
Touchscreens are calibrated from the installed screen, no additional software or computer needed. Note that the instructions are different depending on whether you are using a Wind98 or WinXP Operating system. Instructions for both are shown below
Win98 OS
Exit to OS
Start| Settings| Control Panel| Touch Panel at this point double click on the Touch Panel icon, then select the Calibration Tab.
Check that Use onboard EEProm to store calibration is checked, choose 9 or 20 point calibration and select Com2 to communicate, then touch the Calibration Now button and follow instructions on the screen.
Win XP OS
Exit to OS
Select the PM icon on menu bar. this opens the Pen Mount Control Panel. Select Tab1 Calibrate>Standard Calibration. Follow prompts on screen to complete calibration.