How to Reset Custom Diagnostic Target Reset Option in Essentials 2.0 Follow
Purpose: Provides Reset Option for Custom Diagnostics
Important Notes:
Applies to: ColorFlex L2 Essentials Version 2025.2 and above.
When standardizing in a mode other than the 1" port plate (e.g., Sample Cup or Custom modes), Essentials 2.0 automatically records custom green tile targets for that standardization mode. This occurs the first time you standardize in that mode and present the Check Tile. During the initial standardization with a special port plate, five measurements of the green tile are taken, and their average is set as the target value for future green tile checks in that specific port plate mode.

If the target is incorrectly set, you can reset it in System Menu/Instrument Settings/Diagnostics. After resetting, the next standardization in the affected mode will take five new measurements of the Check Tile to establish a new target value.
Note: This does not apply to the 1" port plate targets, only the Custom and Sample Cup modes are resettable.
Do you need more information? Submit a ticket and a support team member will reach out to you soon!