FAQ: Can I set up multiple transmittance modes with different Top-of-Scale (TOS's) solutions?
Answer:
You can set up to 20 independent standardization modes. For example, you can select SENSOR > SET MODE > ADD MODE. Then change the displayed mode name, Mode #1 to Distilled Water and select type - Total Transmittance (TTRAN). Standardize the mode with a DI Water blank at the Port when prompted to do Top-of-Scale. Next select SENSOR > SET MODE > ADD MODE, then change the displayed mode name, Mode #1 to Basic Solution. Select type as TTRAN, then standardize the mode with a Basic Solution blank at the Port when prompted to read Top-of-Scale. Next select SENSOR > SET MODE > ADD MODE, then change the displayed mode name, Mode #1 to Acid Solution, select type TTRAN, then standardize the mode with an Acid Solution blank at the Port when prompted to read Top-of-Scale.
Each of these standardizations will be stored in memory until the standardization is repeated or 8 hours elapses which ever happens first. If you want to switch modes between readings just click on SENSOR > SET MODE and click on the desired mode, then click on OK.
Do you still need more information? Submit a ticket and our support team member will reach out to you soon!