FAQ: Is there any possibility of averaging the standard in the product setup?
Answer:
Both the MiniScan EZ and ColorFlex EZ create a new cumulative average every time the Product Standard values are created. For example, if you take 4 readings of a physical Product Standard in the setup, the values displayed will be the cumulative average after each reading 1, 2, 3 and 4.
In the MiniScan, you will see the cumulative reading - 1, 2, 3, 4 displayed in the bottom left-hand corner. The ColorFlex EZ does not have this counter but still displays cumulative average of all values read for the Product Standard.
Here's a test to show it. Set up the Averaging feature to average 2 readings.
- Set the Product Standard Values in the setup by first reading the White Tile - L* = 98.77; a* = -3.96; b* = 13.81 for D65/10 illuminant/observe conditions.
- Then read the Green Tile - L* = 60.02; a* = -22.39; b* = 18.26 D65/10.
- After the second reading, the displayed Product Standard values will be L* = 83.14; a* = -8.94; b* = 14.46 D65/10 which is the cumulative average color measurement for both tiles.