FAQ: How to replace the Aeros White Tile.

WarningImportant Notes:
Applies to: Aeros

Answer:

Once a replacement instrument white standard is requested, Hunter Lab will send a new white tile (mounted in a protective centering puck) and a thumbdrive with a csv file to load the new white tile data.

Required: Aeros Essentials 1.03.0074 and above

  1. Attach the thumb drive including the white tile csv file in front USB port of Aeros.
  2. Go to WORKSPACE menu > DIAGNOSTICS > ADVANCED, scroll down the center area and click RESTORE TILE DATA.

 

          3. Select the while tile csv file and click OK to update the white tile TOS data. RESTORE TILE DATA SUCCESSFULLY will be                    shown after updating the TOS data correctly.

 

 

             4. Now, it is ready to use the new white tile to do standardization and measure samples.