FAQ: How can I measure the Haze of a clear plastic bottle?
Answer:
In this case, the ID of the bottle center is 1.75 in diameter; at the threaded mouth, it's 1.76 inches in diameter.
If they will go for making a single wall cut-out (before label applied) from the label panel area, the transmittance C clamp on any of our sphere instruments (USVIS preferred) will work nicely. The single wall corresponds to the protection for the product (look for presence + amount of UV absorber at 370 nm) and also what the consumer sees (CIE L*, a*, b* D65/10 + %Haze, Y Transmittance (C/2 + YI E313 C/2 or D65/10).
If they don't do the cut-out. CMR-2824 Bottle Holder for Custom Transmittance Holder for Transparent Bottles or Cylinders (vertical) as shown in USVIS Measurement Method 5027 should work. Another P# plug would have to be made from their engineering diagram.
Do you still need more information? Submit a ticket and our support team member will reach out to you soon!