Views:

Purpose: How to create a Product Setup Using a Physical Standard & Averaged Multiple Readings and MiniScan EZ.

 

WarningImportant Notes:
Applies to:  MiniScan EZ
  1. Create a Product Setup with a Working Standard and an Average of 3 readings.  Set any other parameters that you require.
  2. Exit to the MAIN MENU and STANDARDIZE the MiniScan EZ.
  3. Select READ and choose the product setup you just created. Since this setup uses a Working standard the first three measurements that you make will be averaged and stored as the Standard measurement. Measure your Standard. View the averaged value of the readings and record them on paper.
  4. EXIT to the Main Menu.
  5. Select PRODUCT SETUP and choose the setup used to measure the Working Standard. Switch the Standard type to PHYSICAL.
  6. Check the Standard Values to ensure that the averaged Standard Measurement is stored, i.e. compare to the written values.