Views:

FAQ: What does each error code mean?

 

WarningImportant Notes:
Applies to:  SpectraTrend

Answer:

Display of codes is cumulative. For example, if 6000 is shown this means both error 4000 and error 2000 occurred. If 1080 is shown this means both error 1000 and error 0080 occurred.

FFFF - photometric data not available

4000 - dark scan fail

2000 signal scan fail

1000 - monitor signal low

0800 - bottom-of-scale signal high

0400 - top-of-scale signal low

0080 - bottom-of-scale not read

0040 - top-of-scale not read

0002 - range too close

0001 - range too far

If photometric data is not available, then the remaining fields are not transmitted.