FAQ: We are currently using a 10mm pathlength and measure 0 - 50 APHA/Pt-Co/Hazen Color, usually in the area of 10. Would you recommend a 10mm pathlength with our equipment for this application?

 

WarningImportant Notes:
Applies to:  UltraScan PRO, UltraScan VIS, VISTA

Answer:

APHA Pt-Co Hazen 0-30-100-200-300-400-500 Liquid Color Standards in Nessler Tubes representing the range of this yellowness scale.

APHA Color is also known by two other names Pt-Co Color and Hazen units (HU). All three names represent the same color scale. As the APHA/Pt-Co/Hazen Color is dependent on the path length or thickness of the sample, it is important to:

  • Use the same cell path length for all measurements.
  • Report the cell path length as part of the measurement method if inter-instrument agreement of APHA/PtCo/Hazen Color values is important at multiple sites.

HunterLab instrument correlation to the visual APHA/Pt-Co/Hazen Color scale per ASTM D5386 allows the user to select a 10-mm, 20-mm or 50-mm path length cell by configuring the Color Data View in EasyMatch QC software to display APHA-10mm, APHA-20mm, or APHA-50mm.

While any of these three cell path lengths can be used to effectively measure APHA/Pt-Co/Hazen Color, the optimal choice for cell path length choice is based on the following criteria:

  • A 10-mm path length cell is typically used for highly saturated or chromatic liquids, typically with APHA/Pt-Co/Hazen Color values > 300.
  • The easy-to-fill 20-mm path length cell is used for moderately chromatic liquids and is the default cell path length when a wide APHA/Pt-Co/Hazen Color range of liquid colors is being measured from 0 (distilled water) to 500.
  • The longer 50-mm path length cell is preferred for clear or near clear liquids where APHA/Pt-Co/Hazen Color values < 30, and especially when APHA < 10.