Purpose: How to connect ColorFlex L2 with EasyMatch QC
Applies to: ColorFlex L2, EasyMatch QC
Note: EZMQC is not recommended for ColorFlex L2. CFL2 operates effectively as a standalone device with Essentials2. Enhanced features in Essentials2 aim to eliminate the need for EZMQC. Additionally, Essentials Desktop software for PC will soon be available for ColorFlex L2.
Accessories Needed: USB Type-A to USB Type-B Cable
- Connect the Cable:
- Plug the USB Type-A end into a USB port on your PC.
- Plug the USB Type-B end into the service port of the CFL2.
- Set Up EZMQC:
- Open the EZMQC software on your PC.
- Navigate to Sensors > Install/Configure.
- Click Add Sensor.
- In the Sensor Type dropdown list, select ColorFlex L2.
- Choose the correct communication port from the list, then click Next.
- Note: If you’re unsure which port corresponds ColorFlex L2, select one and click Next. If it doesn’t work, try another port until successful.
- Verify Connection:
- Once connected, the ColorFlex L2 will be ready for standardization. Restart EZMQC after the sensor is added.
Note: EZMQC cannot communicate with the Essentials2 software on the CFL2 or utilize its camera feature.
Do you need more information? Submit a ticket and a support team member will reach out to you soon!